● Convenience
▶ Conveniently monitored by an independent control display, sterilization process display, and pressure gauge.
- VFD for visualization of temp, pressure(optional), timer, warning alert checks.
- LED lamps for sterilization process check.
- Easy pressure check with the standard gauge attached to the front.
▶ Vertical type autoclaves which allow for top loading are suitable for high volume sterilization.
- Baskets are stackable two high or even more levels, which secure inside the chamber space requirements.
▶ Low-maintenance
- The chamber is constructed of stainless steel which is superiorly resistant to corrosion.
- Easy to replace silicone gasket.
▶ Convenient documentation with the optional data recorder.
- Dot type recorder(6 point) allows for precise documentation.
● Safety
▶ Mechanical safety devices
- Over current protection.
- Over temperature protection by automatic power cutoff.
- Over pressure protection by automatic and manual safety valve.
▶ Safety warning systems
- Over temperature warning.
- Low temperature warning.
- Low water level warning.
- Low heat warning if the temperature does not reach the sterilization or melting temperature.
- Sterilization fail warning if the operation is unexpectedly stopped during sterilization process.
▶ Slide opening door with plastic cover of low thermal conductivity prevents burns from the contact.
Model | ST-50G | ST-65G | ST-85G | ST-105G |
---|
Control | Microprocessor PID control |
Chamber volume (L / cu ft) | 50 / 1.8 | 65 / 2.3 | 85 / 3.0 | 105 / 3.7 |
Temperature 1) | Sterilization temp. (℃ / ℉) | 110 to 123 / 198 to 221.4 |
Melting temp. (℃ / ℉) | 60 to 100 / 108 to 180 |
Pressure | Sterilization at 121℃ (kg/cm2) | 1.1 to 1.2 |
Gauge range in analog (Mpa) | - 0.1 to 0.6 |
Safety | Over temperature protector / Pressure safety valve / Over current protector |
Warning alert (over temp. / low temp. / low heat / sterilization fail / low water level) |
Timer | 0 to 999 min. |
Dimension | No. of baskets (standard / max.) | 2 / 2 | 2 / 2 | 2 / 2 | 2 / 2 |
Load per baskets (Kg / lbs) | 10 / 22.0 |
Basket (ØxH) (mm / inch) | 380x180 / 14.9x7.1 | 380x230 / 14.9x9.1 | 380x330 / 14.9x13.0 | 380x405 / 14.9x15.9 |
Interior (WxH) (mm / inch) | 400x458 / 15.7x18.0 | 400x558 / 15.7x21.9 | 400x758 / 15.7x29.8 | 400x908 / 15.7x35.7 |
Exterior (W×D×H) (mm / inch) | 624x672x870 / 24.6x26.5x34.2 | 624x672x971 / 24.6x26.5x38.2 | 624x672x1083 / 24.6x26.5x42.6 | 624x672x1233 / 24.6x26.5x48.5 |
Exterior with exhaust tank (W×D×H) (mm / inch) | 722x672x870 / 28.4x26.5x34.2 | 722x672x971 / 28.4x26.5x38.2 | 722x672x1083 / 28.4x26.5x42.6 | 722x672x1233 / 28.4x26.5x18.5 |
Net weight (Kg / lbs) | 94.5 / 208.3 | 99.5 / 219.4 | 104 / 229.3 | 110 / 242.5 |
Electrical requirements (230V, 50 / 60 Hz) | 14.3A | 14.3A | 14.3A | 21.3A |
Cat. No. | AAHL1015K | AAHL1025K | AAHL1035K | AAHL1045K |
1) Technical data according to EN 554 / BS EN 285:1997.
* Permissible environmental conditions: temperature (5-40°C) and relative humidity (10-80%).
.png)