Recirculating Coolers




Performance


▶ Temperature range from 3°Cto 40°C.

 High cooling capacity up to 7.1kW at 20°C.

 Powerful circulating pumps up to 40L/min., 4.3 bar.

 Refrigerating system is designed to reduce energy consumption and maximize refrigerating efficiency.

 Pump pressure can be controlled using a by-pass function. (for H type)



● Convenience


 Clear and easy-to-use LCD display. (resolution 0.1°C)

 Bright LED water level indicator can be seen from a distance.

 Built in pressure gauge for checking pump pressure at a glance.

 High/low pressure gauges for condition diagnosis of refrigerating system at a glance. (for H type)

 Wide filling inlet for easy and safe pouring of solution into the unit.

 Suitable model is selectable between magnetic pump, centrifugal pump.

 Double service valve prevents refrigerant leakage for easy maintenance.

 Removable gridded vent for easy maintenance of condenser.

 Built-in casters for easy transport and installation.



● Safety


 Self-diagnostic function identifying errors.

 Complete safety protection system with warning alarms.

- Low fluid level and dry-running protection; when triggered shows an error symbol on the display and an audible buzzer sound.

- Over-temperature protection; If the equipment operates at over maximum temperature due to external environment, user can simply identify the error with a visual signal or an audible buzzer sound.

- Over-current circuit breaker.

 Eco-friendly R-404A refrigerant use for environmental protection.

 Splash-proof keypad.


specifications


Model

HX-20

HX-25

HX-20H

HX-25H

HX-35H

2)HX-45H

2)HX-55H

Temperature1)(bath fluid : water)

Working temperature range (℃ / ℉)

+3 to 40 / +37.4 to 104

Temperaturestability at 15℃ (±℃ / ℉)

1 / 1.8

Cooling capacity(bath fluid : water)

at 20℃ (kW)

1.9

2.4

1.9

2.4

3.6

4.7

7.1

at 10℃ (kW)

1.4

1.7

1.4

1.7

3.1

3.6

5.1

at 5℃ (kW)

1

1.4

1

1.4

2.4

2.9

4.2

Pump

Max. flow rate(L / min, gal / min)

40 / 10.6

28 / 7.39

Max. pressure (bar / psi)

1.43 / 20.74

4.3 / 62.37

Dimension

Max. filling capacity(L / cu ft)

35 / 1.24

45 / 1.59

For tubing dia.(mm / inch)

20 / 3/4

Filling inlet(Ø, mm / inch)

50 / 1.9

Overall (W×L×H, mm / inch)

620×785×990 / 24.4×30.9×39

745×800×1095 / 29.3×31.5×43.1

Net weight (kg / lbs)

93 / 205

101 / 223

97 / 214

105 / 231

120 / 265

130 / 287

140 / 309

Electrical requirements (230V, 60Hz)

8.5A

9A

9.3A

9.8A

15A

19A

Cat. No.

AAH64011K

AAH64021K

AAH64111K

AAH64121K

AAH64131K

AAH64141K

-

Electrical requirements (230V, 50Hz)

7A

7.6A

7.9A

8.8A

14A

Cat. No.

AAH64012K

AAH64022K

AAH64112K

AAH64122K

AAH64132K

-

-

Electrical requirements (120V, 60Hz)

17A

22A

Cat. No.

AAH64013U

AAH64023U

-

-

-

-

-

Electrical requirements (380V, 60Hz)

10A

11.5A

Cat. No.

-

-

-

-

-

AAH64149K

AAH64159K

Electrical requirements (380V, 50Hz)

7A

8.5A

Cat. No.

-

-

-

-

-

AAH64148K

AAH64158K

1) Technical data according to DIN 12876
2) HX-45H, 55H is recorded by 380V, 60Hz.
※ Above specification value is recorded by 50Hz.
※ Product performance may be affected by ambient temperatures